The following image shows some of the different configurations for connecting to Aeroqual Connect or Aeroqual Cloud. Connection methods are set in the Communications column in the Settings area of the Configure app.
Example A. The ethernet mode is set to Auto (DHCP client). In this configuration, the MOXA modem can be used. The modem assigns an IP address to the ePC (192.168.127.10). The WIFI mode is set to Access point so a direct connection to the ePC can be made over local WIFI.
Example B. The ethernet mode is set to Auto (DHCP client). In this configuration, the ePC has been plugged in an office router using an ethernet cable. The router has assigned an IP address to the ePC (192.168.85.177). The WIFI mode is set to Access point so a direct connection to the ePC can be made over local WIFI.
Example C. The ethernet mode is set to Auto (DHCP client). In this configuration, the MOXA modem can be used. The MOXA modem assigns an IP address to the ePC (192.168.127.10). The WIFI mode is set to Client and is connected the office WIFI network. The ePC has been given a WIFI IP address by the router (192.168.0.158). This IP address can be entered directly in the browser to connect to the monitor.
Example D. The ethernet mode is set to Direct (DHCP server). In this configuration, a laptop can be directly connected to the ePC using an ethernet cable. Enter 10.10.0.1 to connect directly to the monitor.