You can change the relay set point for your particular ozone sensor by adjusting the dip switches on the sensor board. If your sensor isn’t listed, contact Aeroqual.
Relay dip switch (1 2 3 4) | Ozone 0-0.150ppm (SM70 UZ) | Ozone 0-0.5ppm (SM70 LZ) | Ozone 0-10ppm (SM70 GZ) |
---|---|---|---|
on on on on | 0.000 | 0.000 | 0 |
off on on on | 0.010 | 0.025 | 0.5 |
on off on on | 0.020 | 0.050 | 1 |
off off on on | 0.030 | 0.075 | 1.5 |
on on off on | 0.040 | 0.100 | 2 |
***off on off on**** | 0.050* | 0.125* | 2.5* |
on off off on | 0.060 | 0.150 | 3 |
off off off on | 0.070 | 0.175 | 3.5 |
on on on off | 0.080 | 0.200 | 4 |
off on on off | 0.090 | 0.250 | 4.5 |
on off on off | 0.100 | 0.225 | 5 |
off off on off | 0.110 | 0.300 | 6 |
on on off off | 0.120 | 0.350 | 7 |
off on off off | 0.130 | 0.400 | 8 |
on off off off | 0.140 | 0.450 | 9 |
off off off off | 0.150 | 0.500 | 10 |
*The relay set point is factory set (unless otherwise specified) to off on off on.