  1. Follow these post-calibration checks and record the results in table 9 in your calibration form. Disconnect your calibration equipment. Disconnect the gas cylinders and safely remove the gas regulators.
    • Follow these post-calibration checks and record the results in table 9 in your calibration form.

    • Disconnect your calibration equipment.

    • Disconnect the gas cylinders and safely remove the gas regulators.

    • Pack away all the calibration gear.

    • Check the gas inlet is secure and do a final inlet flow check.

    • Replace the mesh inlet and cap.

  2. Lock the monitor's door.
    • Lock the monitor's door.

    • Secure the site (if necessary).

    • Exit service mode.

    • Make sure your calibration form is complete.

    • Schedule the next calibration.

  3. Make sure the journal in Aeroqual Connect / Aeroqual Cloud contains all the details of the calibration.
    • Make sure the journal in Aeroqual Connect / Aeroqual Cloud contains all the details of the calibration.

    • The journal automatically records any changes you apply to the offset or gain.

    • You must log when you applied zero air as this is not automatically recorded, unless you are using the AirCal 8000.

    • We recommended you record details about cylinder concentrations.

    • Record which span points you are using and when you started the span gas flowing.

Conclusion

For further support, contact Technical Support.