Introduction
Under some circumstances, sensors may temporarily exhibit a baseline reading higher than zero due to insufficient warm-up, or cross-sensitivity to other gases. In this case it is worthwhile checking against a standard zero air source.
Please allow 3 minutes for sensor warmup, after turning the instrument on and after swapping sensor heads.
Parts
-
-
The calibration accessory has barbed fittings that are intended to fit with 1/8” ID tubing.
-
Care should be taken in selecting a tubing material that is appropriate for the type of gas that you will be using to calibrate the sensor head.
-
PVDF tubing is appropriate for most gases but the tubing's datasheet should be consulted for a full list.
-
Either side of the calibration accessory can be designated the inlet or the outlet.
-
Not all gas that enters through the inlet exits through the calibration accessory outlet; be advised that some gas exits through the sensor head exhaust. Depending on your calibration gas of choice and concentration levels, you may need to perform this calibration under a fume hood.
-
-
-
Make sure the valve on your flow gas regulator is fully off.
-
Screw the gas regulator onto your gas cylinder. When the regulator is correctly fitted, the pressure indicator will show the cylinder pressure.
-
Always point the cylinder and regulator away from yourself and others when attaching or removing a pressure regulator.
-
Secure the cylinder so it can't be knocked over while the regulator is attached.
-
-
-
Connect tubing to gas source and ventilation.
-
Either side of the calibration accessory can be designated the inlet or the outlet
-
-
Step 4 Attach the calibration accessory to the sensor head while the head is plugged into the Ranger
-
Attach the sensor head to Ranger
-
Turn Ranger On
-
Attach the calibration accessory to the sensor head
-
-
-
Check the cylinder pressure to ensure there's enough gas to perform the calibration (10% of full pressure should be enough).
-
Gently turn the regulator valve anticlockwise and flow the target gas at >1L/min until the reading stabilizes (about 10 minutes).
-
-
-
Wait for the gas reading to stabilize on the display of the Ranger
-
-
-
Flow Zero air until the reading stabilizes (approx, 10 minutes).
-
Press and hold the LEFT arrow key to enter the Settings menu. Navigate to Settings > Calibrate
-
Select ‘Re-zero the sensor...’
-
Press OK to confirm.
-
Please wait for the ‘Preparing for calibration’ and ‘Calibration in progress’ screens.
-
Upon successful calibration, the instrument will display a ‘Calibration Complete’ screen.
-
For further support, contact Aeroqual Support.
For further support, contact Aeroqual Support.