Introduction
The PMX should be regularly cleaned to prevent dust accumulation in the event a potentially flammable dust (e.g. coal dust) is measured at high levels. The frequency of cleaning will vary depending on the levels being measured.
This procedure should be carried out in a fume hood or location with adequate ventilation (e.g. outdoors) to prevent inhaling respirable dust.
-
-
Remove the PMX sensor head from the Ranger base
-
Remove the PM inlet from the PMX sensor head by pulling the inlet out of the sensor head.
-
-
-
Attach a piece of 1/8in ID tubing to the barb inside the PMX sensor head.
-
-
-
Blow low pressure compressed air through the PMX sensor head to remove any accumulated dust. Flow for ~10 seconds should be sufficient.
-
-
-
Turn off the compressed air and remove the compressed air tubing from the PMX inlet barb.
-
Reinstall the PMX inlet by pressing the inlet down onto the barb.
-
Reconnect the PMX sensor head to the Ranger base and confirm the sensor is operating correctly.
-