Introduction
Use this guide to manually carry out a span check using the buttons on the front panel of the AirCal 1000.
Tools
No tools specified.
Parts
-
-
Use the UP and DOWN buttons on the front of the AirCal 1000 to select >MENU on the home screen.
-
Press the ENTER button.
-
Use the UP and DOWN buttons to select Manual from the main menu.
-
Press the ENTER button.
-
-
-
Use the UP, DOWN and ENTER buttons to set the port to 1 or 2.
-
-
Use the UP, DOWN and ENTER buttons to set the dilution ratio based on your calculation.
-
Use the DOWN button to select Start and press the ENTER button.
-
For further support, contact Technical Support.