Introduction
The activated carbon and purafil capsules need to be replaced yearly.
A zero air scrubber replacement kit can be purchased from Aeroqual. It comes with all the relevant fittings so it can be installed easily into the AirCal 1000.
Tools
Parts
-
-
Unscrew the 2 top screws at the back of the AirCal 1000. The top plate should slide out easily.
-
-
-
Disconnect the tube between the purafil capsule and the pump.
-
Unscrew the luer fitting connecting the purafil and activated carbon capsules.
-
Unscrew the luer fitting connecting the activated carbon capsule and the heated carulite capsule (not visible).
-
Unclip the purafil and activated carbon capsules and remove.
-
-
-
Clip the new capsules into place.
-
Reconnect the tube between the purafil capsule and pump.
-
Screw the luer fittings to connect the capsules.
-