Introduction
Typically, the sample pump in the particle profiler lasts between 12 and 18 months. However, it should be replaced when the correct flow is no longer being maintained.
To understand how often you should perform this service activity, click here.
Tools
No tools specified.
Parts
-
-
Enter service mode so any fluctuations in the data caused from this activity can be excluded from air quality reports.
-
-
-
Particle profiler pump module contains only 1 x sample pump.
-
The pumps are accessed by removing the pump module assembly. Philips screwdriver can be used to loosen the retaining screws holding the assembly to the back plate of the monitor. Then disconnect and swap out the pump that needs replacing.
-
It is recommended that you check the flow rate after replacing the pump.
-
-
For further support, contact Technical Support.