Introduction
You can use the TSI 440 flowmeter or a single rotameter to measure the flow rate of your particle monitor.
To understand how often you should perform this service activity, click here.
-
-
Enter service mode so any fluctuations in the data caused from this activity can be excluded from air quality reports.
-
-
-
The flow rate of the particle monitor must be 2.0 LPM ± 0.05 (between 1.95 and 2.05 LPM).
-
A constant 2.0 LPM flow is essential to ensure the sharp cut cyclone is separating out the correct particle size for measurement, eg. PM10 or PM2.5.
-
-
-
Remove the TSP head from the top of the sharp cut cyclone and attach the flow adaptor on your flow assembly.
-
Attach the flow assembly to a 0-2.5 LPM volumetric flow meter.
-
If using the R8 flowmeter, connect the assembly to the top port of the flowmeter (negative flow).
-
Ensure the flowmeter is on a steady surface before reading the flow rate.
-
-
-
If the flow isn't correct, adjust the sample flow valve which is located on the face plate of the pump module, until the flow reads 2.0 LPM ± 0.05 (between 1.95 and 2.05 LPM).
-
For further support, contact Technical Support.
For further support, contact Technical Support.