Introduction
The PCX module has a filter which periodically needs to be replaced. Click here for the maintenance schedule.
Tools
Parts
-
-
Enter service mode so any fluctuations in the data caused from this activity can be excluded from air quality reports.
-
-
-
Disconnect power from the PCX module
-
-
-
The filters for the PCX are located inside the module. The front panel has a round filter holder at the bottom. Unscrewing the filter holder reveals the filter.
-
If required, you can use a large flat head screwdriver to unscrew the filter housing on the front panel of the optical module.
-
Replace the filter and reinstall the filter holder, tightening until the filter holder bottoms out.
-
-
-
Reconnect power to the PCX module.
-
-
-
The PCX module can assist you with determining the required service interval as it records the amount of particulate that has passed through the system
-
Click on the ‘Diagnostics and Advanced’ page
-
Click on the ‘Diagnostics' tab
-
The fifth column shows how much PM has passed through the device in mg.
-
-
-
The PCX module has a factory integrated flow sensor will automatically check and maintain the flow rate of the module.
-
For a current reading, navigating to the monitor’s page in Aeroqual Cloud
-
Go to the ‘Diagnostics and Advanced’ page
-
Go to the 'Diagnostics' tab
-
Select the 'Flow' parameter
-
The target flow rate of the PCX module is 1.0 ± 0.05 LPM
-
For further support, contact Technical Support.