Skip to main content

v0.2

Minor Versionm

by Kyle Alberti

Introduction

A leak check should not be necessary in the PCX module. If the system is able to achieve 1L/min at the inlet then measurement accuracy will be maintained.

However, if you find flow rates are not maintained, a leak may have developed.

To understand how often you should perform this service activity, click here.

Tools

No tools specified.

  1. Enter service mode so any fluctuations in the data caused from this activity can be excluded from air quality reports.
    • Enter service mode so any fluctuations in the data caused from this activity can be excluded from air quality reports.

  2. Remove the TSP inlet, remove the two O-rings and place the vacuum gauge on the inlet.
    • Remove the TSP inlet, remove the two O-rings and place the vacuum gauge on the inlet.

    • Wait for the gauge to stabilise.

    • The gauge should reach at least -60 kPa while power is connected.

  3. Stop the sample pump by removing power from the module
    • Stop the sample pump by removing power from the module

    • Count how long it takes for the pressure to change by 10 kPa.

    • If the pressure change (leak rate) is greater than 10 kPa in 10 seconds, you'll need to check for loose connections, worn seals (o-rings) or split tubing.

    • For example, if the needle moves from -70 kPa to -60 kPa in 20 seconds, this is OK, but if if moves from -70 kPa to -60 Kpa in 8 seconds, this indicates a leak.

Conclusion

For further support, contact Technical Support.

Kyle Alberti

Member since: 04/25/2023

9 Guides authored

Team

Aeroqual Staff Member of Aeroqual Staff

2 Members

9 Guides authored