Skip to main content

v0.3

Minor Versionm

by Margaret Grattan

Introduction

Use this procedure to check the zero baseline of your PCX.

The PCX has an automatic zero calibration function which runs at start up and every 24 hour after that, this procedure automatically changes the zero offset of the PCX.

To understand how often you should perform this service activity, click here.

Tools

No tools specified.

  1. Enter service mode so any fluctuations in the data caused from this activity can be excluded from air quality reports.
    • Enter service mode so any fluctuations in the data caused from this activity can be excluded from air quality reports.

  2. Remove the TSP head and O-rings from the top of your monitor’s particle inlet.
    • Remove the TSP head and O-rings from the top of your monitor’s particle inlet.

    • Attach the flow adaptor and zero filter.

  3. To view particle concentrations, open the Diagnostics and Advanced app and select Diagnostics from the side menu. Alternatively, use the Calibration and Service app and select Manual Entry from the side menu.
    • To view particle concentrations, open the Diagnostics and Advanced app and select Diagnostics from the side menu.

    • Alternatively, use the Calibration and Service app and select Manual Entry from the side menu.

    • While the zero filter is attached, the PM µg/m3 values for PCX channels should be 0 or close to 0 (between ±3 µg/m3).

  4. If the zero check shows negative numbers, there may be a problem with the auto zero cycle. See the troubleshooting section.
    • If the zero check shows negative numbers, there may be a problem with the auto zero cycle. See the troubleshooting section.

    • Check that the offset is 0.000

    • Check that the gain is between 0.6 and 4.0

    • The zero baseline check is now complete.

Conclusion

For further support, contact Technical Support.

Kyle Alberti

Member since: 04/25/2023

9 Guides authored